62
N2750/1/2A User’s Guide
6
Performance Verification
Test 1. DC Input Resistance
Refer to the list of required equipment in
NOTE
Allow the probe to warm up for at least 20 minutes.
Procedure
1
Snap the socketed tip onto the probe. The socketed tip is the easiest to use in
this test.
2
Connect the probe to one of the oscilloscope’s channel inputs. This ensures
that the probe is powered.
3
Turn on the oscilloscope.
4
Using the digital multimeter, measure the DC input resistance between leads
A
and
B
. This is the differential resistance and should measure 200 k
Ω
± 2%.
5
Measure the single-ended input resistance between lead
A
and the ground
lead. The resistance should measure 100 k
Ω
± 2%.
6
Measure the single-ended input resistance between lead
B
and the ground
lead. The resistance should measure 100 k
Ω
± 2%.
7
Enter the measured values in
Summary of Contents for N2750A