Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
39
8
Wait until the measurement settles.
9
Press the
Y2
softkey, then position the Y2 cursor to the center of the voltage
trace using the Entry knob.
The
Δ
Y value on the lower line of the display should be within the test limits of
(depending on the oscilloscope model).
If a result is not within the test limits, go to the “Troubleshooting” chapter. Then
return here.
10
Disconnect the precision source from the oscilloscope.
11
Repeat this procedure to check the dual cursor accuracy with the remaining
Volts/div setting values in
(depending on the oscilloscope
model).
12
Finally, repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise
may be a factor. To eliminate the noise, add a BNC Tee, blocking capacitor, and
shorting cap at the oscilloscope channel input to shunt the noise to ground. See
. If a BNC capacitor is not available, use an SMA blocking capacitor,
adapter, and cap. See
for
details.
Figure 6
Using a Blocking Capacitor to Reduce Noise
To oscilloscope input
BNC shorting
cap
Blocking
Capacitor
Summary of Contents for InfiniiVision 2000
Page 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Page 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Page 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Page 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...