E2655C Probe Deskew and Performance Verification Kit User’s Guide
Running the Probe Calibration and Deskew
1
On the Infiniium oscilloscope in the Setup menu, select the channel connected
to the probe.
2
In the Channel Setup dialog box, select the
Probe Config...
button.
3
Select the Probe head model from the drop-down menu.
4
In the Probe Configuration dialog box, select the
Calibrate Probe...
button.
5
In the Probe Calibration dialog box, select the
Start Atten/Offset Calibration...
button and follow the on-screen instructions for the vertical calibration
procedure.
6
Once the vertical calibration has successfully completed, select the
Start Skew
Calibration...
button and follow the on-screen instructions for the skew
calibration.
At the end of each calibration the oscilloscope will inform you if the calibration
was or was not successful.
Figure 4
Bad Connection