E2655C Probe Deskew and Performance Verification Kit User’s Guide

13

 

10

Select Setup menu and choose the channel connected to the BNC cable from 

the pull-down menu.

11

Select the 

Probes...

 button.

12

Select the 

Configure Probe System

 button.

13

Select 

User Defined Probe

 from the pull-down menu.

14

Select the 

Calibrate Probe...

 button.

15

Once the skew calibration is completed, close all dialog boxes.

16

Select the 

Start Skew Calibration...

 button and follow the on-screen 

instructions.

17

Set the vertical scale for the displayed channels to 

100 mV/div

.

18

Set the horizontal range to 

1.00 ns/div

.

19

Set the horizontal position to approximately 

3 ns

.

20

Change the vertical position knobs of both channels until the waveforms 

overlap each other.

21

Select the Setup menu choose 

Acquisition...

 from the pull-down menu.

22

In the Acquisition Setup dialog box enable averaging. When you close the 

dialog box, you should see waveforms similar to that in 

Figure 6

.

Figure 6

Overlapping Waveforms

Summary of Contents for E2655C

Page 1: ...Keysight E2655C Probe Deskew and Performance Verification Kit User Guide...

Page 2: ...ides the Software to U S gov ernment customers under its standard com mercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www key sight com fi...

Page 3: ...E2655C Probe Deskew and Performance Verification Kit User s Guide 3...

Page 4: ...4 E2655C Probe Deskew and Performance Verification Kit User s Guide...

Page 5: ...1131B 1132B 1134B and the InfiniiMode differential probes N2750A N2751A N2752A is done using the E2655C Probe Deskew and Performance Verification Kit The kit contains the following parts SMA male to S...

Page 6: ...tion for the InfiniiMax 1160B series and 1130B series probes solder in differential probe head and the differential browser probe head However the procedures also apply to all of the different Infinii...

Page 7: ...nfiniium oscilloscope 4 Connect the probe to an oscilloscope channel 5 To minimize the wear and tear on the probe head the probe head should be placed on a support to relieve the strain on the probe h...

Page 8: ...uide NOTE To ensure contact pull up on the back side of the pincher to ensure good contact between resistor leads and the deskew fixture Figure 2 Connecting the Probe for Calibration BNC to SMA Connec...

Page 9: ...front panel 2 Set the volts per division to 100 mV div 3 Set the horizontal scale to 1 00 ns div 4 Set the horizontal position to approximately 3 ns You should see a waveform similar to that in Figur...

Page 10: ...he Probe Configuration dialog box select the Calibrate Probe button 5 In the Probe Calibration dialog box select the Start Atten Offset Calibration button and follow the on screen instructions for the...

Page 11: ...ctor farthest from the pincher 3 Connect the BNC male to BNC male cable to the BNC connector on the deskew fixture to one of the unused oscilloscope channels For Infiniium oscilloscopes with bandwidth...

Page 12: ...e Auto Scale key on the front panel NOTE To ensure contact pull up on the back side of the pincher to ensure good contact between resistor leads and the deskew fixture Figure 5 Connection setup for ve...

Page 13: ...boxes 16 Select the Start Skew Calibration button and follow the on screen instructions 17 Set the vertical scale for the displayed channels to 100 mV div 18 Set the horizontal range to 1 00 ns div 1...

Page 14: ...the browser on the center conductor of the deskew fixture between the green line and front end of the pincher The negative resistor tip or ground pin of the browser must be on either of the two outsi...

Page 15: ...E2655C Probe Deskew and Performance Verification Kit User s Guide 15 Figure 7 Calibration setup BNC to SMA Connector not visible InfiniiMax Probe Pincher Deskew Fixture 50 W SMA Terminator...

Page 16: ...16 E2655C Probe Deskew and Performance Verification Kit User s Guide...

Page 17: ...E2655C Probe Deskew and Performance Verification Kit User s Guide 17...

Page 18: ...This information is subject to change without notice Keysight Technologies 2003 2004 2013 2016 Edition 5 August 2016 E2655 92004 www keysight com...

Reviews: