E2655C Probe Deskew and Performance Verification Kit User’s Guide
9
Verifying the Connection
1
On the Infiniium oscilloscope, press the
[Auto Scale]
key on the front panel.
2
Set the volts per division to
100 mV/div
.
3
Set the horizontal scale to
1.00 ns/div
.
4
Set the horizontal position to approximately
3 ns
. You should see a waveform
similar to that in
.
If you see a waveform similar to that of
then you have a bad connection
and should check all of your probe connections.
Figure 3
Good Connection
Summary of Contents for E2655C
Page 1: ...Keysight E2655C Probe Deskew and Performance Verification Kit User Guide...
Page 3: ...E2655C Probe Deskew and Performance Verification Kit User s Guide 3...
Page 4: ...4 E2655C Probe Deskew and Performance Verification Kit User s Guide...
Page 16: ...16 E2655C Probe Deskew and Performance Verification Kit User s Guide...
Page 17: ...E2655C Probe Deskew and Performance Verification Kit User s Guide 17...