8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
95
Measurement Operations & Settings
4
example:
sens1:corr:coll:zero
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also zeroed.
command:
:SENSe[
n
]:[CHANnel[
m
]]:CORRection:COLLect:ZERO?
syntax:
:SENSe[
n
]:[CHANnel[
m
]]:CORREction:COLLect:ZERO?
description:
Returns the status of the most recent zero command.
parameters:
none
response:
0:
97
98
99
any other number:
zero succeeded without errors.
zero not done
zero failed
zero absorbed
remote zeroing failed (the number is the error code returned from the operation).
example:
sens1:corr:coll:zero?
->
0<END>
affects:
All power meters and return loss modules
dual sensors:
Master and slave channels are independent.
command:
:SENSe[
n
]:[CHANnel[
m
]]:CORRection:COLLect:ZERO:ALL
syntax:
SENSe[
n
]:[CHANnel[
m
]]:CORRection:COLLect:ZERO:ALL
description:
Zeros the electrical offsets for all installed power meter and return loss modules.
parameters:
none
response:
none
example:
sens:chan:corr:coll:zero:all
affects:
All power meters and return loss modules
dual sensors:
Command is independent of channel.