8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
55
Instrument Setup and Status
3
Status Reporting
–
The STATus Subsystem
The Status subsystem allows you to return and set details from the Status
Model. For more details, see
command:
:STATus:OPERation[:EVENt][:LEVel
0
]?
syntax:
:STATus:OPERation[:EVENt][:LEVel
0
]?
description:
Returns the Operational Status Event Summary Register (OSESR).
parameters:
none
response:
The sum of the results for the slots (a
16-bit signed integer
value, where 0
≤
value
≤
32767):
Bits
Mnemonics
Decimal Value
8163A/B
8164A/B
8166A/B
15
Not used
Not used
Not used
0
14
Not used
Not used
Slot 14 Summary
16384
13
Not used
Not used
Slot 13 Summary
8192
12
Not used
Not used
Slot 12 Summary
4096
11
Not used
Not used
Slot 11 Summary
2048
10
Not used
Not used
Slot 10 Summary
1024
9
Not used
Not used
Slot 9 Summary
512
8
Not used
Not used
Slot 8 Summary
256
7
Not used
Not used
Slot 7 Summary
128
6
Not used
Not used
Slot 6 Summary
64
5
Not used
Not used
Slot 5 Summary
32
4
Not used
Slot 4 Summary Slot
4
Summary 16
3
Not used
Slot 3 Summary
Slot 3 Summary
8
2
Slot 2 Summary
Slot 2 Summary
Slot 2 Summary
4
1
Slot 1 Summary
Slot 1 Summary
Slot 1 Summary
2
0
Not used
Slot 0 Summary
Level 1 Summary
1
example:
stat:oper?
->
+0<END>