8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
187
Measurement Operations & Settings
4
command:
:OUTPut
n
[:CHANnel[
m
]]:APMode?
syntax:
:OUTPut
n
[:CHANnel[
m
]]:APMode?
description:
Queries whether the user has amended the power value or the attenuation value.
This use of this command is optional.
Included for compatibility with Keysight 8156A mainframe.
parameters:
none
response:
boolean
0 User has amended the attenuation value.
1 User has amended the power value.
example:
OUTP1:APMode?
->
0<END>
affects:
All attenuator modules.
command:
:OUTPut
n
[:CHANnel[
m
]]:POWer
syntax:
:OUTPut
n
[:CHANnel[
m
]]:POWer<wsp><value>[PW | NW | UW | MW | W | DBM ] | MIN | MAX | DEF
description:
Sets the output power value (
P
).
If your attenuator module does not include the power control feature, the new filter attenuation (
a
filter (new)
) is calculated
from the reference power (
P
ref
) :
P
set(new)
(dBm) = P
ref
(dBm) -
a
filter (new)
(dB) -
P
Offset
(dB)
If your attenuator module includes the power control feature, the filter attenuation is changed until the set power (measured
by the module’s internal power meter) has been reached.
P
set(new)
(dBm) = P
att(new)
(dBm) - P
offset
(dB)
If the set power cannot be achieved ExP (indicating ’Excessive Power’ ) is displayed, and the appropriate GPIB status bit is set.
The status of these bits can be queried using
:STATus:OPERation:CONDition[:LEVel0]?
parameters:
<value>
MIN | MAX | DEF
Desired output power (if unit not specified current unit is used).
Sets the output power to the module limits, or the module default.
response:
none
example:
OUTP1:POW 12
affects:
All attenuator modules.
command:
:OUTPut
n
[:CHANnel[
m
]]:POWer?
syntax:
:OUTPut
n
[:CHANnel[
m
]]:POWer<wsp>MIN | MAX | DEF
description:
Without the optional parameter, queries the output power value.
parameters:
MIN | MAX | DEF
Queries the output power module limits, or the module default.