84
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4
Measurement Operations & Settings
command:
:SLOT[
n
]:HEAD[
m
]:WAVelength:RESPonse?
syntax:
:SLOT[
n
]:HEAD[
m
]:WAVelength:RESPonse?
description:
Returns the wavelength response from a wavelength calibrated module in binary format.
response:
Wavelength Response table as a
binary block
.
response
format:
One 8 byte long wavelength calibration value pair consisting of a 4 byte long float for wavelength and a 4 byte long float for
the scalar calibration factor.
For more information on binary block formats see
example:
slot1:head1:wav:resp?
->
#536570........
affects:
Attenuator with power control, all powermeters, return loss modules