294
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
9
Error Codes
Old
-213
“Init ignored”
Description:
The INIT:IMM command (
) initiates a trigger and completes a full measurement
cycle. The continuous measurement must be DISABLED. This error code is generated if
the powermeter is still in cont. measurement mode.
Old
-220
"Parameter error (StatParmError)”
Description:
The user has passed a parameter that cannot be changed in this way. The device cannot
detect one of the following more specific errors:
Old
-220
-220, "Parameter error (StatParmOutOfRange)"
Description:
The user has passed a parameter that exceeds the valid range for this parameter.
Old
-220
"Parameter error (StatParmIllegalVal)"
Description:
The user has passed a parameter that does not match a value in a list of possible values.
New/Old/Standard
Error
Number
String