Performance Tests
1
Keysight 53210A/53220A/53230A Assembly Level Service Guide
65
Test 6a: (Q) Frequency accuracy - channel 2/3, option 106
(6 GHz microwave channel)
This test verifies the frequency accuracy of Option 106 Channel 2 (53210A) or
Channel 3 (53220A/53230A) at two frequencies (100 MHz and 5.88 GHz).
Equipment
– FS725 Rubidium timebase
– Keysight E8257D-520 microwave signal generator
– Keysight 10503A BNC cable (2)
– Keysight 11500E (3.5) mm cable
– Keysight 1250-1250 N(M) to SMA(F)
E8257D-520 setup
– Output Frequency:
100 MHz
– Output Power:
-27 dBm
NOTE
The input power MUST be -27 ±0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 100 MHz or a power meter
to verify input power.