– +907, “Self-test failed; PIC communication failure”
– (1) 53200-66501 Motherboard; (2) Battery; (3) P0500-63001 processor
board
– +908, “Self-test failed; Battery Failed”
– (1) 1420-0909 battery; (2) 53200-66507 battery charger assy
– +909, “Self-test failed; GPIB test Failed”
– (1) P0001-20001 GPIB board; (2) P0500-63001 processor board
– +910, “Self-test failed; Channel 3 test failed”
– (1) 53200-66505 (Option 106) or 53200-66506 (Option 115);
– (2) 532xx-66501 motherboard
– +911, “Self-test failed; Front Panel revision check failed”
– (1) 53200-66502 front panel; (2) P0500-63001 processor board
– +912, “Self-test failed; Measurement board test failed”
– (1) 53200-66501 Motherboard; (2) P0500-63001 processor board