
2
Service
124
Keysight 53210A/53220A/53230A Assembly Level Service Guide
8
If “
Lock Cal
” is displayed on a softkey, the counter is unlocked and ready for
calibration.
– The
Perform Cal
softkey becomes available.
9
Press the
Perform Cal
softkey:
– The
Cal Step
softkey shows the last cal step that was selected:
10
Press the
Cal Step
softkey.
– Display softkey annunciators show:
* Only displayed if Option 106 or Option 115 is installed.
11
Press the
Oscil
softkey.
– Display softkey annunciators show:
*
NOTE
The data entry area will display the expected calibration frequency (10.000
MHz).