Service
2
Keysight 53210A/53220A/53230A Assembly Level Service Guide
143
10
If “
Lock Cal
” is displayed on a softkey, the counter is unlocked and ready for
calibration.
– The
Perform Cal
softkey becomes available.
11
Press the Perform Cal softkey:
– Display softkey annunciators show:
12
Press the
Cal Step
softkey.
– Display softkey annunciators show:
* Only displayed if Option 106 or 115 is installed.
13
Press the
uWave
softkey.
– Display softkey annunciators show:
14
Press the
Perform Zero-Cal
softkey to perform zero-calibration.
– Display momentarily shows:
**
NOTE
If your microwave source does not have an ‘output enable/disable’ feature,
temporarily disconnect the cable from the microwave input channel until after
the zero-calibration is performed.