1
Performance Tests
42
Keysight 53210A/53220A/53230A Assembly Level Service Guide
13
Press the
Statistics
softkey:
– The
Statistics
softkeys are displayed:
14
Press to set:
– Statistics:
On
– The display shows:
Absolute timebase test procedure
1
Press the
Trigger
key to manually trigger the counter.
– Counter takes 100 readings and fills in Math functions.
2
Verify the ‘Mean’ displays a reading of:
– 10 MHz ± 10 Hz (for standard timebase)
– 10 MHz ± 0.5 Hz (for Option 010 high stability timebase)
3
Record the ‘Mean’ test value, Test 1, in the
“Keysight 53210A/53220A/53230A
Performance Test Record (Tests 1 to 7)”
4
Proceed to Test #2 or disconnect the test setup.