DosiPump 1000 User Manual
Page 18 / 24
Version 1.01E
In order to replace the device fuse, use a small screwdriver to
lever the fuse holder of the power receptacle down and out.
Fuse:
3.15 A - 250 V AC - slow-blow
No fuses with other ratings may be used!
WARNING
Always disconnect the mains cable first before replacing the fuse!
Faults and errors occurring during operation
Symptom
Possible cause
Correction
Fault signal
Pump blocked
The culture medium in the tube is too
viscous. You have used a tube which
is too large.
Remove the tube and run the pump
manually with the "PUMP" menu.
Fault signal:
No program entered
The selected program has no name
(has not been edited).
Edit the selected program and assign
it a program name.
Fault signal:
DosiPump barred
No valid license
License has expired
Apply for a new license from the sell-
er with the license application code
and enter the license code.