2 Safety information
2.1 Operating environment
Several factors must be noted when selecting the right operating environment for the unit. The room se-
lected may not pose an aggressive environment either for the unit itself or for its intended application.
The D
OSI
P
UMP
1000 should be operated only under ambient conditions corresponding to a normal envi-
ronment (5°C with maximum 80 % relative humidity).
NOTE
These conditions also apply to transport.
The D
OSI
P
UMP
1000 should be installed on a clean, dry, smooth, horizontal surface. The front panel of
the unit should be near to the front edge of the workbench in order to ensure convenient operation and
easy access to the touch screen. The unit should be installed in the direct vicinity of electrical socket
outlets.
Summary of the D
OSI
P
UMP
1000 operating environment
Temperature range:
+5°C to +45°C
Relative humidity:
Max. 80 % up to 31°C; decreasing in linear fashion down to 50 % at 40°C
Mains power supply:
Voltage fluctu/- 10 % of rated value
Pollution severity:
2 in accordance with IEC 664
Location:
Complying with enclosure IP 21
Sturdy support surface
Environment:
Avoid splash water and contact with chemicals.
DosiPump is not an EX Approved
WARNING
We are unable to guarantee operability and functional safety of the unit if
the ambient conditions applicable to the location are not complied with.
2.2 Safety-conscious working
The D
OSI
P
UMP
1000 complies with the state of the art and is safe to operate. The D
OSI
P
UMP
1000 may
pose residual risks and dangers if it is used and operated incorrectly by untrained personnel. All persons
entrusted with the task of operating the D
OSI
P
UMP
1000 must have read and understood the User Manu-
al and, in particular, the safety information or must have been instructed by superiors in such a manner
that they are able to safely operate the D
OSI
P
UMP
1000.
Besides the safety information listed above, other applicable provisions and regulations such as GLP,
GMP, FDA, the regulations of the German Liability Insurance Associations, those of the Heath Authority
and those of the Factory Inspectorate must also be followed.
2.3 Qualification of the operating personnel
Specialist technical training is generally required. Special knowledge is required when operating the
D
OSI
P
UMP
1000 to prevent potentially dangerous situations with the unit. Non skilled staff without the
corresponding specialist technical training generally do not have such qualifications and are able to per-
form these tasks only after instruction and under the supervision of specialist personnel.