5 ISO/GLP-compliant Printout or Record
Application:
Use of the balance as a test and measuring instrument in quality assurance systems in
compliance with the requirements of ISO,GLP,GMP and EN (European Standards) in
which proof of the balance’s accurate performance and adjusting is required.
The balance can record all completed adjusting and adjustment operations and print out
data in compliance with the requirements of Good Laboratory Practice (GLP). The
balance, interfaced with a data printer or a computer, creates a document that records
the date, time, serial number and model number, making it possible to clearly trace data
to the balance that generated it and the point of time at which it was generated.
Select the ISO/GLP-compliant printout or record mode by setting the respective
code in the balance operating menu:
ISO/GLP-compliant printout or record mode
Code
Off
8 10 1*
Only for adjusting and linearisation functions
8 10 2
Always on (e.g. adjusting and linearisation functions, weight readouts)
8 10 3
The following menu code setting must be selected in order to obtain an ISO/GLP-
compliant printout/record:
with data ID code
7 2 2
Important Note !
If the factory setting - code 7 2 1 - is selected, ISO/GLP-compliant printouts or records
will not be generated.
In addition, do not select the “Auto print“ data output parameter (code 6 1 4 or 6 1 5).
Operating the balance with a Device that has an ISO/GLP Printing or Recording
Capability.
To the
KERN 770, GS
and
GJ
balances a special Data printer can be connected for
generating ISO/GLP-compliant printouts.
This printer provides you with the following functions:
- Turns the ISO/GLP function on and off;
- Date/Time;
- ID no. for identifying a workstation/operator;
- Printouts with balance-specific data.