![KERN 770-12 Operating Instructions Manual Download Page 42](http://html1.mh-extra.com/html/kern/770-12/770-12_operating-instructions-manual_1957145042.webp)
770/GS/GJ-BA-defsi-0023 109
4.9 Practical Example “Animal Weighing in the Manual Start Mode“
Menu code settings used in this example:
Function
Code
Animal Weighing
2 1 7
Average Start delay
3 7 2*
Manual Start mode
3 8 1
The number of sub-weighing operations is changeable.
If you do not enter a number, averaging will be done based on the number shown on the
display (factory setting: 10 sub-weighing operations).
Simple weighing of animals
Balance Readout
Place animal weighing bowl on balance.
CF
,
TARE
0.00
g
Place 1st animal in bowl
press
F
10
432.41
g
Remove 1st animal
Place 2nd animal in bowl
press
F
10
391.53
g
Remove 2nd animal
Repeat above steps for all animals.
* = factory setting