Technical Data
75
Technical Data
Features specification
Ambient conditions
-
indoor use
-
maximum elevation 2000 m (6562 ft) above sea level
-
max. relative humidity 80 % up to 31°C (88°F), linearly decreasing
down to 50 % relative humidity at 40°C (104°F).
Ambient temperature allowed
+2 °C to +35 °C (36°F to 95°F)
run time
- standard mode :
1min – 9 h 59 min, hold = permanent
- extended mode :
1min – 99 h, hold
maximum speed (n
max
)
15 000 rpm
(rotor-dependent, adjustable in steps of 10)
minimum speed (n
min
) 300
rpm
maximum RCF
21 885
(micro liter rotor)
maximum kinetic energy Legend™ T
Legend™ RT
45.8 kNm
(33 780 ft.lb)
54.8 kNm
(40 418 ft.lb)
noise at maximum speed Legend™ T
Legend™ RT
< 66 dB (A)
< 55 dB (A)
(sound pressure level of emission according to DIN EN ISO 11 201)
Temperature set range
-9 °C to +40 °C