Technical Data
77
Electrical connections / fuses
Order no.
Voltage
Frequency Nominal
current
Power
consumption
Fuse protection of
instrument
-
safety fuse
* thermic excess current
release
Fuse
protection of
building
Legend™ T
7500 4366
120 V
60 Hz
10.2 A
900 W
15 A
15 AT
Legend™ T
7500 4362
230 V
50/60 Hz
8.6 A
1250 W
10 A
16 AT
Legend™ RT
7500 4376
120 V
60 Hz
12.0 A
1220 W
15 A
*
16
A
15 AT
Legend™ RT
7500 4372
230 V
50/60 Hz
10.6 A
1900 W
10 A
*
16
A
16 AT