Operation
42
Installing the rotor
Improper or improperly combined
accessories may cause severe dam-
age to the centrifuge!
The rotors
approved for the
Legend™
are detailed in
the chapter "Rotors and accessories". Use only rotors
listed for this instrument.
To install the rotor
,
you need the socket wrench sup-
plied* (see chapter "Accessories").
* The swinging buckets rotor 7500 6434 comes com-
plete with a locking mechanism.
Proceed as follows:
1. Open the lid and make sure that the rotor chamber
is clean. Clean any dust, foreign material or sample
residues out of chamber before use.
2. Check whether the collet chuck is loose (collet
chuck moves freely on the spindle). If not, loosen
the rotor seat using the socket wrench supplied.
3. Place the rotor on top of the collet chuck so that the
rotor is located precisely above the center.
4. The rotor must glide freely down the collet chuck
until it hits the lower stop.
5. If you have positioned the rotor correctly, you can
tighten the collet chuck easily using the hexagon
wrench supplied.
6. Place the rotor cover on applicable rotors and
tighten securely.
Regularly check the proper positioning
of the rotor and re-tighten the collet
chuck as needed.
Please take care of the readability of the inscription of
the swinging bucket rotor cross installed.
(rotor identification must show to the chamber bottom)
For the swinging bucket rotor the set bucket type must
be permitted for the operation in the respective rotor.
Please note the hints of chapter "Bucket selection of
swinging bucket rotors"
Summary of Contents for Sorvall Legend RT
Page 1: ...SORVALL Legend T RT Instruction Manual QUIKset...
Page 10: ...For your safety 6 Notes...
Page 16: ...Description of the 12 Notes...
Page 44: ...Rotors and accessories 40 Notes...
Page 60: ...Operation 56 Notes...
Page 66: ...Maintenance and care 62 Notes...
Page 82: ...78 Notes...
Page 88: ...84...
Page 100: ...96...