Rotors and accessories
20
Table 1: Rotors for Legend™
T
100V / 120V (differences of 230V instruments are shown in parentheses)
Rotor designation
Order no.
Fixed-Angle Rotor
6 x 94 ml
7500 3334
High
conic
®
Rotor
6 x 50 ml
7500 3057
Maximum permissible load [ g ]
6 x 140
6 x 130
Maximum speed n
max
[ rpm ]
10 500
8500
Maximum RCF value at n
max
12 449
10 016
Radius (max. / min.) [ cm ]
10.1 / 6.3
12.4 / 6.0
Tube angle [ ° ]
26
45
Acceleration / deceleration time [ s ]
45 / 45
( 40 / 45 )
33 / 35
( 25 / 35 )
Heating of samples at n
max
[ °C ]
relative to room temperature 23°C,
Run time 1 hour
13°C
10°C
Aerosol-tight *
yes
yes
Autoclavable
no
121°
* Checked by CAMR, Porton-Down, UK