Rotors and accessories
37
BIOshield
®
-Rotor
Highplate
®
-Rotor
The aerosol-tight bio-containment of air
vessel rotors is only warrented in a
horizontal position!
If the air vessel rotors are filled with samples outside
the centrifuge (e.g. in a safety work bench), one has to
care, that during the installation into the centrifuge the
rotor should be kept in a vertical position.
Closing the aerosol-tight bio-containment
After greasing the seal, turn the cover until it sits lightly
on the bucket.
To achieve uniform seal, turn the cover clockwise by
1½ grip area ( approx. 15°). Use the setting marks on
the bucket as a guide.
Bucket mark
Grip area