11.  Prompt examinations for some parameters 

Press [Fn] button to examine some parameters in normal working. Every time press 
it, contents of one item will be displayed. If it keeps on some contents more than 6 
seconds, it will return to normal working state automatically. 
 
Content “1PXXXXX” is displayed first, it is comparative value of 1

relay action. 

And then content “2PXXXXX”is comparative value of 2

relay action. “ALXXXXX”is 

corresponding weight value of bottom of analog output, and “AHXXXXX” is 
corresponding weight value of top of analog output. 
 

12.  Error indicating information 

E1: overload of excitation loop or short circuit of load cell 
E2: restriction for button operation, operating weight calibration under hardware 
protection 
E3: restriction for button operation, change mode of analog output under hardware 
protection  
E4: parameters are incorrect when calibrating, uV value of each division is less 
than 0.6uV. 
E5: restriction for button operation, adjust bottom and top of analog output under 
hardware protection   

 

13.  Maintenance of indicator 
13.1. Regular  maintenance 

Tiny maintenance is needed. In clean environment, maintenance for one time each 
year is enough. But in dirty environment, times for maintenance should be 
increased accordingly. Soft cloth can be used to clean the board. Any industrial 
solvent and scour must not be used to spray indicator. Meanwhile, pay attention to 
dust and water in daily use. 
 
If interior circuitry board needs cleaning, power supply must be turn off. After that 
open the outside box, and blow dust off from the board with dry high-pressure air.     

13.2. Malfunction resolve

 

Phenomenon 1: Nothing is displayed after transducer is electrified.   
Solution:    1. Check power supply 
     

    2. Check fuse 

 
Phenomenon 2: Indicator displays “E1” 
Solution:  1. Check load cell plug if it is short circuit   

2. Check signal wire and junction box cable if they are damaged or 

short circuit 

 

Phenomenon 3: After Self Test, indicator displays “

―――――――

”all the time 

Solution:  1. Check load cell plug if its connection is well 

2. Check voltage if it is in the right range 

Summary of Contents for XK3101

Page 1: ...XK3101 KM05 Weighing Transducer User Manual z Read the manual carefully before use z Please take good care of this manual for future reference www kelichina com Page 1 of 17...

Page 2: ...converter maximum A D pulse 1 000 000 z Updating rate 6 25 times second 12 5 times second 25 times second 50 times second z Number of its division 1000 10000 z Division value range 0 001 50 z Two poin...

Page 3: ...nd it cannot share power supply with electrical equipments for example relay or heater which easily causes noise d Temperature and Humidity Working temperature 20 50 humidity 10 95 no condensation Sav...

Page 4: ...fixed to both sides of weighing terminal tightly to guarantee the firm installation of weighing transducer 2 2 Connection of weighing transducer a Back View picture of weighing transducer Fig 8 4 3 b...

Page 5: ...lectric welding or other strong electric operations are prohibited on scales Action to prevent damage caused to load cell and indicator from thunder must be taken in raining season Safety for both ope...

Page 6: ...ple connecting higher wire of current to 1 end lower wire to 2 end As voltage output positive part connects to 3 end and negative part to 4 end Transducer cannot output voltage and current signals sim...

Page 7: ...r light POWER power supply indicator light OVER overload of load cell excitation loop or short circuit indicator light Dynamic it is bright when weight reading is unstable Tare tare indicator light Ze...

Page 8: ...lue choosing a suitable divisional value must guarantee that uV of each division is more than or equal to 0 6uV d Calculation of uV d is as following formula Divisional value kg sensitivity of load ce...

Page 9: ...urn to next step automatically g When indicator displays calibrate the weighing value Standard poise is put on the scale and the poise whose weight is more than 20 of maximum weighing value is recomme...

Page 10: ...ess Fn button it can switch to display When it displays press PT button it means accepting and saving the calibrated result When it displays press PT button it means the calibrated result is available...

Page 11: ...s Fn button for three times continuously indicator displays c Press PT button for two times continuously indicator press Fn button to choose parameter F4 2 0 weight transducing range is from zero to m...

Page 12: ...of analog output top Fine adjustment of analog output top Elaborate adjustment of analog output top Under adjusting items press 0 button to increase value or press Tare button to reduce value the req...

Page 13: ...lly 0 prohibition 1 allowance auto zero tracking 0 5d second 2 allowance auto zero tracking 1d second 3 allowance auto zero tracking 3d second F2 5 dynamic testing 0 prohibit dynamic testing 1 allowan...

Page 14: ...ss 0 button to change value e Press PT button indicator saves out 2 and displays F5 1 X set mode of relay output Press Fn button to choose parameters 0 no action of relay 1 weight separate selective m...

Page 15: ...mer s request 9 2 Choose assemble ways Serial interface receive data by two ways RS232 and RS485 and change wire position on circuitry board and choose one Open transducer backboard pull out circuitry...

Page 16: ...6uV E5 restriction for button operation adjust bottom and top of analog output under hardware protection 13 Maintenance of indicator 13 1 Regular maintenance Tiny maintenance is needed In clean envir...

Page 17: ...ontinuous transmit mode Phenomenon 6 No action of rely Solution 1 Check comparative value of out 1 and out 2 2 Check working mode of relay not work weight separate selection or fixed mode Index Calibr...

Reviews: