Note: 

when uV of each division is less than 0.6 uV, indicator will show “E4”, 

then press [ Tare ] button, it returns to the previous step. Press [ PT ] button, the 

calibrated result can be accepted. In other words, uV of each division is less 

than 0.6nV, indicator still work correctly. 

k.  Press [ PT ] button, indicator displays “

”, meanwhile the light 

pole is extinguished gradually. During this time the reading is unstable and the 

light pole comes back to the full-bright state. After light pole fully extinguished, 

indicator will turn to next step automatically. 

l.  The value of loaded poise should be inputted when indicator displays maximum 

weighing value XXXXX. Press [ Fn ] button, position of glittering can be chosen 

and press [

0

] button, value of glittering bits can be amended.   

m. Press [ PT ] button, indicator displays “

”, then press [ Fn ] button, it 

can switch to display “

”. When it displays “

”, press 

[ PT ] button, it means accepting and saving the calibrated result. When it 

displays “

”, press [ PT ] button, it means the calibrated result is 

available only this time, when turn on the transducer again, it adopts the 

previous calibrated data. Until now, calibration is finished.     

 

Tips: 

If the zero drift of load cell exceeds the range of zero set-ups when power on, 

Zero can be calibrated again as calibration steps. When it is step 7, press 

[ Tare ] button for many times, and when indicator displays “

”, press [ PT ] 

button to quit out.   

Notice: incorrect operation – such as press [ PT ] button when displaying 

”, it will turn into capacity-proofread step. So you must be careful 

when calibrating zero separately.   

 

6.  Choice and adjustment of analog output modes 

Calibrating head must be inserted and then the calibration can be operated. It is 

only available when indicator possesses analog output module. 

6.1. 

Choose and adjustment of output mode 

www.kelichina.com       

                                                                                                                            Page 10 of 17 

Summary of Contents for XK3101

Page 1: ...XK3101 KM05 Weighing Transducer User Manual z Read the manual carefully before use z Please take good care of this manual for future reference www kelichina com Page 1 of 17...

Page 2: ...converter maximum A D pulse 1 000 000 z Updating rate 6 25 times second 12 5 times second 25 times second 50 times second z Number of its division 1000 10000 z Division value range 0 001 50 z Two poin...

Page 3: ...nd it cannot share power supply with electrical equipments for example relay or heater which easily causes noise d Temperature and Humidity Working temperature 20 50 humidity 10 95 no condensation Sav...

Page 4: ...fixed to both sides of weighing terminal tightly to guarantee the firm installation of weighing transducer 2 2 Connection of weighing transducer a Back View picture of weighing transducer Fig 8 4 3 b...

Page 5: ...lectric welding or other strong electric operations are prohibited on scales Action to prevent damage caused to load cell and indicator from thunder must be taken in raining season Safety for both ope...

Page 6: ...ple connecting higher wire of current to 1 end lower wire to 2 end As voltage output positive part connects to 3 end and negative part to 4 end Transducer cannot output voltage and current signals sim...

Page 7: ...r light POWER power supply indicator light OVER overload of load cell excitation loop or short circuit indicator light Dynamic it is bright when weight reading is unstable Tare tare indicator light Ze...

Page 8: ...lue choosing a suitable divisional value must guarantee that uV of each division is more than or equal to 0 6uV d Calculation of uV d is as following formula Divisional value kg sensitivity of load ce...

Page 9: ...urn to next step automatically g When indicator displays calibrate the weighing value Standard poise is put on the scale and the poise whose weight is more than 20 of maximum weighing value is recomme...

Page 10: ...ess Fn button it can switch to display When it displays press PT button it means accepting and saving the calibrated result When it displays press PT button it means the calibrated result is available...

Page 11: ...s Fn button for three times continuously indicator displays c Press PT button for two times continuously indicator press Fn button to choose parameter F4 2 0 weight transducing range is from zero to m...

Page 12: ...of analog output top Fine adjustment of analog output top Elaborate adjustment of analog output top Under adjusting items press 0 button to increase value or press Tare button to reduce value the req...

Page 13: ...lly 0 prohibition 1 allowance auto zero tracking 0 5d second 2 allowance auto zero tracking 1d second 3 allowance auto zero tracking 3d second F2 5 dynamic testing 0 prohibit dynamic testing 1 allowan...

Page 14: ...ss 0 button to change value e Press PT button indicator saves out 2 and displays F5 1 X set mode of relay output Press Fn button to choose parameters 0 no action of relay 1 weight separate selective m...

Page 15: ...mer s request 9 2 Choose assemble ways Serial interface receive data by two ways RS232 and RS485 and change wire position on circuitry board and choose one Open transducer backboard pull out circuitry...

Page 16: ...6uV E5 restriction for button operation adjust bottom and top of analog output under hardware protection 13 Maintenance of indicator 13 1 Regular maintenance Tiny maintenance is needed In clean envir...

Page 17: ...ontinuous transmit mode Phenomenon 6 No action of rely Solution 1 Check comparative value of out 1 and out 2 2 Check working mode of relay not work weight separate selection or fixed mode Index Calibr...

Reviews: