select parameter by pressing [Fn] button. Press [PT] button to next parameter
choice.
d. Press [Tare] button for many times at any time, Press [PT] button to quit out
when indicator displays “
ESC
”
7.2.
Content of function F2 choice parameter
F2.1 to choose ADC transmitting speed rate
0
=
7.5Hz
;
1=15Hz
;
2=30Hz
;
3=60Hz (available when turn it on next time)
F2.2 to tare
0=prohibition, 1=allowance tare range 100%FS
F2.3 to clear to zero
0=prohibition
1=allowance clear zero range ±4%FS
;
2=allowance clear zero range ±10%FS
;
3=allowance clear zero range ±20%FS
;
F2.4 set zero tracking range automatically
0=prohibition
1=allowance auto zero tracking 0.5d/second
2=allowance auto zero tracking 1d/second
3=allowance auto zero tracking 3d/second
F2.5 dynamic testing
0=prohibit dynamic testing
1=allowance dynamic testing sensitivity 0.5d
2=allowance dynamic testing sensitivity 1d
3=allowance dynamic testing sensitivity 3d
F2.6 numeric filter choice
Number stands for intensity of filter, number is the bigger, filter is stronger, stable
time becomes longer accordingly
F2.7 auto zero-set range when turn it on
1=auto zero-set range ±4%FS when turn it on
2=auto zero-set range ±10%FS when turn it on
3=auto zero-set range ±20%FS when turn it on
8. Relay
output
In-built two-point relay output of transducer, action mode of relay can be set: 1 no
action of relay, 2 weight separate selective mode, 3 fixed value mode