output will work according to chosen weight transducing range. 

 

6.3. 

Adjust bottom and top of analog output 

Calibrating four analog outputs before delivery, thus choosing mode of analog 
output is enough when using it. Bottom and top of analog output can be changed if 
necessary, for example, you can set analog output range from 1V to 4.5V. 
 
Testing adjusting value with high-precise voltage or current meter, or connect to the 
former machine to adjust it directly as well. 

 

Adjustment steps: 

 

a.  Press [Fn] and [Tare] buttons simultaneously, indicator displays “

”. 

b.  Press [PT] button continuously, indicating prompt of adjusting items can be 

displayed circularly. 

  : Coarse adjustment of analog output bottom 

  : Fine adjustment of analog output bottom 

:    Elaborate adjustment of analog output bottom 

Coarse adjustment of analog output top 

Fine adjustment of analog output top 

Elaborate adjustment of analog output top 

:  

Under adjusting items, press [0] button to increase value or press [Tare] button to 
reduce value the required one and then press [Fn] button to quit out. 
 
Adjustment just can affect current set-up analog output mode. 

 

6.4. 

Reset of bottom and top of analog output. 

Setting bottom and top of analog output as original value, when users adjust it to be 
disordered, it can be reset quickly. 
 
Reset method: 
a.  Press [Fn] and [Tare] buttons simultaneously, indicator displays “

 

b.  Press [Fn] button continuously, indicator change to “

” and “

 

c.  Press [PT] button when displaying “

”, turn it on again and the 

analog output will be original analog quickly. 

 

7.  Indicator parameter selection

 

7.1. 

Turn into working parameters choice 

a.  Press [Fn] and [0] buttons simultaneously, indicator displays “

” 

b.  Press [Fn] button, indicator displays “

” 

c.  Press [PT] button, turn into parameter choice “F2.1”, each parameter choice to 

www.kelichina.com       

                                                                                                                            Page 12 of 17 

Summary of Contents for XK3101

Page 1: ...XK3101 KM05 Weighing Transducer User Manual z Read the manual carefully before use z Please take good care of this manual for future reference www kelichina com Page 1 of 17...

Page 2: ...converter maximum A D pulse 1 000 000 z Updating rate 6 25 times second 12 5 times second 25 times second 50 times second z Number of its division 1000 10000 z Division value range 0 001 50 z Two poin...

Page 3: ...nd it cannot share power supply with electrical equipments for example relay or heater which easily causes noise d Temperature and Humidity Working temperature 20 50 humidity 10 95 no condensation Sav...

Page 4: ...fixed to both sides of weighing terminal tightly to guarantee the firm installation of weighing transducer 2 2 Connection of weighing transducer a Back View picture of weighing transducer Fig 8 4 3 b...

Page 5: ...lectric welding or other strong electric operations are prohibited on scales Action to prevent damage caused to load cell and indicator from thunder must be taken in raining season Safety for both ope...

Page 6: ...ple connecting higher wire of current to 1 end lower wire to 2 end As voltage output positive part connects to 3 end and negative part to 4 end Transducer cannot output voltage and current signals sim...

Page 7: ...r light POWER power supply indicator light OVER overload of load cell excitation loop or short circuit indicator light Dynamic it is bright when weight reading is unstable Tare tare indicator light Ze...

Page 8: ...lue choosing a suitable divisional value must guarantee that uV of each division is more than or equal to 0 6uV d Calculation of uV d is as following formula Divisional value kg sensitivity of load ce...

Page 9: ...urn to next step automatically g When indicator displays calibrate the weighing value Standard poise is put on the scale and the poise whose weight is more than 20 of maximum weighing value is recomme...

Page 10: ...ess Fn button it can switch to display When it displays press PT button it means accepting and saving the calibrated result When it displays press PT button it means the calibrated result is available...

Page 11: ...s Fn button for three times continuously indicator displays c Press PT button for two times continuously indicator press Fn button to choose parameter F4 2 0 weight transducing range is from zero to m...

Page 12: ...of analog output top Fine adjustment of analog output top Elaborate adjustment of analog output top Under adjusting items press 0 button to increase value or press Tare button to reduce value the req...

Page 13: ...lly 0 prohibition 1 allowance auto zero tracking 0 5d second 2 allowance auto zero tracking 1d second 3 allowance auto zero tracking 3d second F2 5 dynamic testing 0 prohibit dynamic testing 1 allowan...

Page 14: ...ss 0 button to change value e Press PT button indicator saves out 2 and displays F5 1 X set mode of relay output Press Fn button to choose parameters 0 no action of relay 1 weight separate selective m...

Page 15: ...mer s request 9 2 Choose assemble ways Serial interface receive data by two ways RS232 and RS485 and change wire position on circuitry board and choose one Open transducer backboard pull out circuitry...

Page 16: ...6uV E5 restriction for button operation adjust bottom and top of analog output under hardware protection 13 Maintenance of indicator 13 1 Regular maintenance Tiny maintenance is needed In clean envir...

Page 17: ...ontinuous transmit mode Phenomenon 6 No action of rely Solution 1 Check comparative value of out 1 and out 2 2 Check working mode of relay not work weight separate selection or fixed mode Index Calibr...

Reviews: