9-4
System Expansion (TSP-Link)
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Front panel operation
The TSP-Link node number can be set from the MAIN MENU of each instrument;
this process is summarized in
.
Table 9-1
Assigning a node number to an instrument from the front panel
Series 2600 MAIN MENU
1) Press the
MENU
key to access
MAIN MENU.
2) Select the
COMMUNICATION
menu.
3) Select the
TSPLINK_CFG
menu.
4) Select the
NODE
menu.
5) Set the
NODE
number (
01
to
64
) for the instrument.
Remote programming
The
tsplink.node
attribute is used to set the node number for an instrument:
tsplink.node = N
where
: N = 1 to 64
The node number of an instrument can be determined by reading the
tsplink.node
attribute as follows:
print(tsplink.node)
The above
command will output the node number. For example, if the node
number is 1, the value
1.000000e00
will be output.
Resetting the TSP-Link
After all the node numbers are set, you must initialize the system by performing a
TSP-Link reset. For initialization to succeed, all units must be powered on when
the TSP-Link reset is performed.
NOTE
If you change the system topology after
initialization, you must re-initialize the system by
performing a TSP-Link reset. Changes that affect the
system topology include:
• Powering down or rebooting any unit in the sys-
tem.
• Rearranging or disconnecting the LAN cable
connections between units.