![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2470/sourcemeter-2470_user-manual_4042495067.webp)
Model 2470 High Voltage SourceMeter Instrument User's Manual
Section 6: Measure I-V characteristics of FETs
2470-900-01 Rev. A /
May
2019
6-9
Send the following commands for this example application:
SMU 1,
SMU 2, or
pseudocode
Commands
Description
SMU 1
*RST
:SENS:FUNC "CURR"
:SENS:CURR:RANG:AUTO ON
:ROUT:TERM REAR
:SOUR:FUNC VOLT
:SOUR:VOLT:RANG 20
:SOUR:CONF:LIST:CRE "stepVals"
:DIG:LINE1:MODE TRIG, OUT
:DIG:LINE2:MODE TRIG, IN
:TRIG:DIG1:OUT:STIM NOT1
:TRIG:DIG2:IN:CLE
:TRIG:DIG2:IN:EDGE RIS
Reset the instrument.
Set to measure current.
Set to measure with autorange
enabled.
Set to rear terminals.
Set to source voltage.
Set to 20 V source range.
Create a source configuration list
called
stepVals
.
Set digital line 1 to trigger out.
Set digital line 2 to trigger in.
Set the stimulus for digital line 1 to
the notify1 event.
Clear digital line 2.
Set to detect rising edge on line 2.
Pseudocode
for i = 2 to 5 do:
Set up a
for
loop from 2 to 5.
SMU 1
:
SOUR:VOLT i
:
SOUR:CONF:LIST:STORE "stepVals"
Set the voltage level to the iteration
number.
Store the source configuration to
stepVals
.
Pseudocode
end for
End the
for
loop.
SMU 1
:
TRIG:BLOC:CONF:RECALL 1, "stepVals"
:
TRIG:BLOC:SOUR:STAT 2, ON
:
TRIG:BLOC:MEAS 3
:
TRIG:BLOC:NOT 4, 1
:
TRIG:BLOC:WAIT 5, DIG2
:
TRIG:BLOC:CONF:NEXT 6, "stepVals"
:
TRIG:BLOC:BRAN:COUN 7, 4, 3
:
TRIG:BLOC:SOUR:STAT 8, OFF
Create trigger model block 1 to load
stepVals
at the first index.
Create block 2 to turn the output on.
Create block 3 to make a
measurement.
Create block 4 to generate the
notify1
event.
Create block 5 to wait on digital line
2.
Create block 6 to load the next
index of
stepVals
.
Create block 7 to branch to block 3,
3 times.
Create block 8 to turn the output off.
Summary of Contents for SourceMeter 2470
Page 2: ...High Voltage SourceMeter Instrument User s Manual Model 2470...
Page 6: ......
Page 24: ......
Page 44: ......