![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2470/sourcemeter-2470_user-manual_4042495047.webp)
Model 2470 High Voltage SourceMeter Instrument User's Manual
Section 5: Leakage current and insulation resistance
2470-900-01 Rev. A /
May
2019
5-3
Figure 21: Capacitor leakage and insulation resistance test connection schematics
The following figures show the rear-terminal connections to the device under test (DUT) for these
applications. If capacitor leakage measurements are noisy, you may need to use the high
capacitance mode or add a low leakage forward-biased diode in series with the capacitor.
Figure 22: Rear-panel connections for the leakage current test
Figure 23: Rear-panel connections insulation resistance test