Command Reference
B-9
Detecting calibration errors
If an error occurs during any calibration step, the Model 2400 will generate an appropriate
error message. Several methods to detect calibration errors are discussed in the following
paragraphs.
Reading the error queue
As with other Model 2400 errors, any calibration errors will be reported in the error queue.
(You can read the error queue by using the :SYST:ERR? query.)
Error summary
Table B-4 summarizes calibration errors. These errors will occur under the following
conditions.
Error 500/501
Will occur if you do not set the calibration date or calibration due date.
Error 502
Will occur when the :CAL:PROT:SENS or :CAL:PROT:SOUR command is
sent with a parameter that cannot be placed into any of the defined numeric
ranges.
Error 503/504
Will occur when the source/compliance DAC calibration constants generate
an invalid DAC setting. All of the source/compliance DAC calibration
constants are verified each time a :CAL:PROT:SOUR command is
processed and when the Model 2400 is first turned on. This error generally
occurs when the Model 2400 source circuitry is defective.
Error 505
Will occur when a :CAL:PROT:SOUR command generates an invalid
source offset calibration constant.
Error 506
Will occur when a :CAL:PROT:SOUR command generates an invalid
source gain calibration constant.
Error 507
Will occur when a :CAL:PROT:SENS command generates an invalid
measurement offset calibration constant.
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......