Model 2400 Service Manual
2-2
Calibration
2400-902-01 (G - Feb 2006)
Introduction
Use the procedures in this section to calibrate the Model 2400. These procedures require
accurate test equipment to measure precise DC voltages and currents. Calibration can be
performed either from the front panel or by sending SCPI calibration commands over the
IEEE-488 bus or RS-232 port with the aid of a computer.
WARNING
The information in this section is intended for qualified service personnel
only. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages.
Environmental conditions
Temperature and relative humidity
Conduct the calibration procedures at an ambient temperature of 18-28•C (65-82•F) with
relative humidity of less than 70% unless otherwise noted.
Warm-up period
Allow the Model 2400 to warm up for at least one hour before performing calibration.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument’s internal temperature to stabilize. Typically,
allow one extra hour to stabilize a unit that is 10•C (18•F) outside the specified temperature
range.
Also, allow the test equipment to warm up for the minimum time specified by the
manufacturer.
Line power
The Model 2400 requires a line voltage of 100 to 240V at line frequency of 50 or 60Hz. The
instrument must be calibrated within this range.
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......