Keithley Instruments, Inc.
Cleveland, Ohio 44139
DRN.
CKD.
APP.
DATE
DATE
DATE
PART NUMBER
LTR
REVISIONS
APP. DATE
FORM 28777A-SBG
BRUNING 40-21 62198-SBG
SPECIFICATIONS
SPEC-2440
A 25977
Released
SZ 5/22/01
B 27103
Revised
SZ 5/20/02
C 28003
Revised
SZ 3/17/03
HW 5/22/01
SZ 5/22/01
HW 3/26/03
Rev. C
Page 1 of 3
2440 and 2440-C SourceMeter
®
Specifications
SOURCE SPECIFICATIONS
VOLTAGE PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
PROGRAMMING
RESOLUTION
ACCURACY (1 Year)
23°C
±
5°C
±
(% rdg. + volts)
NOISE
(peak-peak)
0.1Hz – 10Hz
200.000 mV
5 µV
0.02% + 600 µV
10 µV
2.00000 V
50 µV
0.02% + 600 µV
50 µV
10.0000 V
500 µV
0.02% + 1.2 mV
250 µV
40.0000 V
5 mV
0.02% + 4.8 mV
1 mV
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy specification)/
°C.
MAX. OUTPUT POWER:
55W, four quadrant source or sink operation.
SOURCE/SINK LIMITS:
±
10.5V @
±
5.25A,
±
42V @
±
1.05A.
VOLTAGE REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 100
µV
.
NOISE 10Hz – 1MHz (p-p):
50mV typical.
OVERVOLTAGE PROTECTION:
User selectable values, 5% tolerance. Factory default = none.
CURRENT LIMIT:
Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT:
<0.1% typical (full scale step, resistive load, 10mA range).
CURRENT PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
PROGRAMMING
RESOLUTION
ACCURACY (1 Year)
2
23°C
±
5°C
±
(% rdg. + amps)
NOISE
(peak-peak)
0.1Hz – 10Hz
10.0000 µA
500 pA
0.033% + 2 nA
50 pA
100.000 µA
5 nA
0.031% + 20 nA
500 pA
1.00000 mA
50 nA
0.034% + 200 nA
5 nA
10.0000 mA
500 nA
0.045% + 2 µA
50 nA
100.000 mA
5 µA
0.066% + 20 µA
500 nA
1.00000
1
A
50 µA
0.067% + 900 µA
50 µA
5.00000
1
A
50 µA
0.10% + 5.4 mA
500 µA
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy specification)/
°C.
MAX. OUTPUT POWER:
55W, four quadrant source or sink operation.
SOURCE/SINK LIMITS:
±
5.25A @
±
10.5V,
±
1.05A @
±
42V.
CURRENT REGULATION: Line:
0.01% of range.
Load:
0.01% (5 A range 0.05%) of range +
100
pA
.
VOLTAGE LIMIT:
Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT:
<0.1% typical (1mA step, RL = 10k
Ω
, 10V range).
ADDITIONAL SOURCE
SPECIFICATIONS
TRANSIENT RESPONSE TIME:
30
µs minimum for the
output to recover to its spec. following a step change in load
.
COMMAND PROCESSING TIME:
Maximum time
required for the output to begin to change following the
receipt of :SOURce:VOLTage|CURRent <nrf> command.
Autorange On:
10ms.
Autorange Off:
7ms.
OUTPUT SETTLING TIME:
Time required to reach 0.1%
of final value after command is processed. 100µs typical.
Resistive load. 10µA to 100mA range.
OUTPUT SLEW RATE (±30%):
0.25V/µs, 40V range, 100mA compliance.
0.08V/µs, 10V range, 100mA compliance.
DC FLOATING VOLTAGE:
Output can be floated up to
±40VDC from chassis ground.
REMOTE SENSE:
Up to 1V drop per load lead.
COMPLIANCE ACCURACY:
Add 0.3% of range and
±0.02% of reading to base specification.
OVER TEMPERATURE PROTECTION:
Internally
sensed temperature overload puts unit in standby mode.
RANGE CHANGE OVERSHOOT:
Overshoot into a fully
resistive 100k
Ω
load, 10Hz to 1MHz BW, adjacent range
changes between 200mV, 2V, and 10V ranges, 100mV
typical.
MINIMUM COMPLIANCE VALUE:
0.1% of range.
1. Full power source operation regardless of load to 30°C ambient. For
above 30°C and/or power sink operation, refer to the Power
Equations section of the User’s Manual.
2. For sink mode, 10µA to 100mA range, accuracy is:
±(0.5% + offset*3)
For 1A and 5A range, accuracy is:
±(1.5% + offset*3)
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......