Keithley Instruments, Inc.
Cleveland, Ohio 44139
DRN.
CKD.
APP.
DATE
DATE
DATE
PART NUMBER
LTR
REVISIONS
APP. DATE
FORM 28777A-SBG
BRUNING 40-21 62198-SBG
SPECIFICATIONS
SPEC-2410
A 19554
Released
SZ 2/26/97
B 19738
Revised
SZ 3/24/97
C 22669
Revised
SZ 5/11/99
D 27953
Revised
SZ 3/10/03
SJ 1-31-97
CC 2/21/97
HW 3/20/03
Rev. D
Page 1 of 3
2410 and 2410-C SourceMeter
®
Specifications
SOURCE SPECIFICATIONS
1
VOLTAGE PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
PROGRAMMING
RESOLUTION
ACCURACY (1 Year)
23°C
±
5°C
±
(% rdg. + volts)
NOISE
(peak-peak)
0.1Hz – 10Hz
200.00 mV
5 µV
0.02% + 600 µV
5 µV
2.00000 V
50 µV
0.02% + 600 µV
50 µV
20.0000 V
500 µV
0.02% + 2.4 mV
5 mV
1000.00 V
50 mV
0.02% + 100 mV
20 mV
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy specification)/
°C.
MAX. OUTPUT POWER:
22W, four quadrant source or sink operation.
SOURCE/SINK LIMITS:
±
21V @
±
1.05A,
±
1100V @
±
21mA.
VOLTAGE REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 1m
V
.
NOISE 10Hz – 1MHz (p-p):
20mV typical into a resistive load.
OVER VOLTAGE PROTECTION:
User selectable values, 5% tolerance. Factory
default = none.
CURRENT LIMIT:
Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT:
<0.1% typical (full scale step, resistive load, 20mA range).
CURRENT PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
PROGRAMMING
RESOLUTION
ACCURACY (1 Year)
3
23°C
±
5°C
±
(% rdg. + amps)
NOISE
(peak-peak)
0.1Hz – 10Hz
1.00000 µA
50 pA
0.035% + 600 pA
5 pA
10.0000 µA
500 pA
0.033% + 2 nA
50 pA
100.000 µA
5 nA
0.031% + 20 nA
500 pA
1.00000 mA
50 nA
0.034% + 200 nA
5 nA
20.0000 mA
500 nA
0.045% + 4 µA
200 nA
100.000 mA
5 µA
0.066% + 20 µA
1 µA
1.00000 A
2
50 µA
0.27% + 900 µA
100 µA
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy specification)/
°C.
MAX. OUTPUT POWER:
22W, four quadrant source or sink operation.
SOURCE/SINK LIMITS:
±
1.05A @
±
21V,
±
21 mA @
±
1100V.
CURRENT REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 1n
A
.
VOLTAGE LIMIT:
Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT:
<0.1% typical (1mA step, RL = 10k
Ω
, 20V range).
ADDITIONAL SOURCE
SPECIFICATIONS
TRANSIENT RESPONSE TIME:
30
µs minimum for the
output to recover to its spec. following a step change in load
.
COMMAND PROCESSING TIME:
Maximum time
required for the output to begin to change following the
receipt of :SOURce:VOLTage|CURRent <nrf> command.
Autorange On:
10ms.
Autorange Off:
7ms.
OUTPUT SETTLING TIME:
Time required to reach 0.1%
of final value after command is processed. 100µs typical.
Resistive load. 10µA to 100mA range.
OUTPUT SLEW RATE (±30%):
0.5V/µs, 1000V range, 20mA compliance.
0.15V/µs, 20V range, 100mA compliance.
DC FLOATING VOLTAGE:
Output can be floated up to
±250VDC from chassis ground.
REMOTE SENSE:
Up to 1V drop per load lead.
COMPLIANCE ACCURACY:
Add 0.3% of range and
±0.02% of reading to base specification.
OVER TEMPERATURE PROTECTION:
Internally
sensed temperature overload puts unit in standby mode.
RANGE CHANGE OVERSHOOT:
Overshoot into a fully
resistive 100k
Ω
load, 10Hz to 1MHz BW, adjacent range
changes between 200mV, 2V, and 20V ranges, 100mV
typical.
MINIMUM COMPLIANCE VALUE:
0.1% of range.
1. Specifications valid for continuous output currents below 105mA.
For operation above 105mA continuous for > 1 minute, derate
accuracy 10%/35mA above 105mA.
2. Full operation (1A) regardless of load to 30°C. Above 30°C ambient,
derate 35mA/°C and prorate 35mA/
Ω
load. 4-wire mode. For current
sink operation on 1A range, maximum continuous power is limited
to approximately 1/2 rated power or less, depending on current, up to
30°C ambient. See power equations in the User’s Manual to
calculate allowable duty cycle for specific conditions.
3. For sink mode, 1µA to 100mA range, accuracy is:
±(0.5% + offset*3)
For 1A range, accuracy is:
±(1.5% + offset*3)
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......