B-10
Command Reference
Error 508
Will occur when a :CAL:PROT:SENS command generates an invalid
measurement gain calibration constant.
Table B-4
Calibration errors
Error number
Error message
+500
“Date of calibration not set”
+501
“Next date of calibration not set”
+502
“Calibration data invalid”
+503
“DAC calibration overflow”
+504
“DAC calibration underflow”
+505
“Source offset data invalid”
+506
“Source gain data invalid”
+507
“Measurement offset data invalid”
+508
“Measurement gain data invalid”
Status byte EAV (Error Available) bit
Whenever an error is available in the error queue, the EAV (Error Available) bit (bit 2) of the
status byte will be set. Use the *STB? query to obtain the status byte, then test bit 2 to see if it
is set. If the EAV bit is set, an error has occurred, and you can use the appropriate error query to
read the error and at the same time clear the EAV bit in the status byte.
Generating an SRQ on error
To program the instrument to generate an IEEE-488 bus SRQ (Service Request) when an error
occurs, send the *SRE 4 command. This command will enable SRQ when the EAV bit is set.
You can then read the status byte and error queue as outlined above to check for errors and to
determine the exact nature of the error.
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......