6-12
Troubleshooting
3. If the accessory slots are functional, use the functional board to check
the I/O connections. Reconnect and check the operation of the I/O
connections, one at a time.
4. If operation fails for an I/O connection, check the individual inputs
one at a time for shorts and opens.
5. If operation remains normal to this point, the problem is in the
PDISO-8 board(s) originally in the computer. If you were using more
than one board, try each board one at a time in the computer to
determine which is faulty.
6. If you cannot isolate the problem, refer to the next section for
instructions on obtaining assistance.
Technical Support
Before returning any equipment for repair, call the Keithley Hardware
Applications Engineering Department at:
1-888-KEITHLEY
Monday - Friday, 8:00 a.m. - 5:00 p.m., Eastern Time
An applications engineer will help you diagnose and resolve your
problem over the telephone. Please make sure that you have the following
information available before you call: