Model 6517A Getting Started Manual
Measurement Options
2-23
Figure 2-6
Combining math calculations
Test sequences
The following information summarizes the built-in test sequences of the Model 6517A.
Detailed information on test sequences can be found in the User’s Manual.
Selecting and configuring a test sequence
A test sequence is selected and configured from the CONFIGURE SEQUENCE menu as
follows:
1.
Press CONFIG and then SEQ to display the sequence configuration menu.
2.
Place the cursor (using
and
keys) on APPLICATIONS and press ENTER.
Select and configure a test from the test sequence menu structure. To select a menu
item, place the cursor (using
and
keys) on it and press ENTER. To change a
parameter value, use the cursor and RANGE keys, and press ENTER.
DEV-CHAR — Use to select a device characterization test:
DIODE: Diode leakage current test — Specify start voltage, stop voltage, step
voltage and delay.
CAPACITOR: Capacitor leakage current test — Specify bias voltage, bias
time, and discharge time.
CABLE: Cable insulation resistance test — Specify bias voltage and bias time.
RESISTOR: Resistor voltage coefficient test — Specify source 1 voltage,
delay 1, source 2 voltage, and delay 2.
RESISTIVITY — Use to select a resistivity test:
SURFACE: Surface resistivity test — Specify pre-discharge time, bias volt-
age, bias time, measure voltage, measure time, and discharge time.
VOLUME: Volume resistivity time — Specify pre-discharge time, bias volt-
age, bias time, measure voltage, measure time, and discharge time.
Math
Polynomial,
Percent,
Percent Deviation,
Ratio, Logarithmic
Value Displayed
Raw
Measurement
Multiple Displays
MAX, MIN