Performance Verification
5-7
Figure 5-3
Connections for ohms verification
* Includes Model 263 error.
COMMON
MODEL 263
BA
OUTPUT
Table 5-3
Limits for ohms verification
6512
range
263 output
(nominal)
6512 & 263
guard
Calculated limit
Allowable 6512 reading
(1 year, 18°-28°C)
263 reading
Equipment
error*
Limit
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
200M
Ω
2G
Ω
20G
Ω
200G
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
10M
Ω
100M
Ω
1G
Ω
10G
Ω
100G
Ω
Off
Off
Off
Off
Off
Off
On
On
On
(____
×
0.16%)
+ 0.0004k
Ω
= ____
(____
×
0.13%)
+ 0.001k
Ω
= ____
(____
×
0.23%)
+ 0.01k
Ω
= ____
(____
×
0.225%)
+ 0.0001M
Ω
= ____
(____
×
0.2125%) + 0.001M
Ω
= ____
(____
×
0.23%)
+ 0.01M
Ω
= ____
(____
×
1.4%)
+ 0.0001G
Ω
= ____
(____
×
1.275%)
+ 0.001G
Ω
= ____
(____
×
1.1%)
+ 0.01G
Ω
= ____
263 Reading ± Limit
263 Reading ± Limit
263 Reading ± Limit
263 Reading ± Limit
263 Reading ± Limit
263 Reading ± Limit
263 Reading ± Limit
263 Reading ± Limit
263 Reading ± Limit