![background image](http://html2.mh-extra.com/html/keithley/6487/6487_user-manual_3951658043.webp)
2-8
Connections
Interlock
The Model 6487 has a built-in interlock that works in conjunction with the voltage source.
The interlock prevents the voltage source from being placed in operate on the 50V and
500V ranges, and optionally on the 10V range, to assure safe operation.
Interlock connections
shows interlock connections and the pin diagram of the INTERLOCK connec-
tor. Typically, the INTERLOCK connector is connected to the same type of connector on
the test fixture. A normally open switch is connected to pins 1 and 2 of the INTERLOCK
connector as shown. When the switch is open, the interlock is asserted, and the voltage
source cannot be placed in operate on the 50V or 500V voltage source ranges and option-
ally for the 10V range.
WARNING
If the voltage source was operating when the interlock is asserted, the
voltage source will change to a high impedance state, possibly leaving
charged DUT capacitance.
Interlock operation
The interlock is always operational for the 50V and 500V voltage source ranges. To enable
the voltage source output, pins 1 and 2 of the INTERLOCK connector must be shorted
together. For the 10V range, the interlock is optional and can be controlled with instru-
ment programming. See
“Voltage source operation,” page 3-11
programming the interlock operation for the 10V range.
Summary of Contents for 6487
Page 13: ......
Page 15: ......
Page 35: ...1 20 Getting Started Model 6487 User s Manual...
Page 47: ...2 12 Connections Model 6487 User s Manual...
Page 61: ...3 14 Measurements and Sourcing Voltage Model 6487 User s Manual...
Page 85: ...7 8 Remote Operation and Commands Model 6487 User s Manual...
Page 86: ...A Specifications...
Page 88: ...B GeneralMeasurement Considerations...
Page 94: ...C ExamplePrograms...
Page 101: ......
Page 103: ......
Page 104: ......