Connections
2-3
WARNING
The maximum safe voltage between the voltage source or ammeter
common and chassis ground (common mode voltage) is 505V peak.
Exceeding this voltage can create a shock hazard.
Figure 2-2
Maximum input levels
Connecting cables and test leads
When making precision measurements, you should always use low-noise triax cables for
INPUT connections. As a general rule, always use the shortest possible cable for measure-
ments. When using the voltage source, the test leads must be rated for 505V minimum and
should include safety sheaths. See
“Input cables, connectors, and adapters,” page 1-4
for recommended cables and test leads.
Input HI
Input LO
505V Peak
Chassis Ground
Max Continuous
Input = 505V Peak
505V Peak