Remote Operation and Commands
7-3
Figure 7-1
IEEE-488 and RS-232 connector locations
Front panel GPIB operation
Error and status messages
See
in the Model 6487 Reference Manual for a list of error and status
messages associated with IEEE-488 programming.
GPIB status indicators
REM shows when the instrument is in the remote state, TALK is on when the instrument is
in the talker active state, and LSTN is on when the Model 6487 is in the listener active
state. SRQ shows when the instrument has generated a service request.
LOCAL key
The LOCAL key cancels the remote state and restores local operation of the instrument.
Pressing the LOCAL key also turns off the REM indicator and returns the display to nor-
mal if a user-defined message was displayed. If the unit is in LOCAL (not in remote), the
LOCAL key acts as a configure key (see
“Front panel summary,” page 1-7
If the LLO (Local Lockout) command is in effect, the LOCAL key is also inoperative.
120
LINE RATING
50, 60Hz
50 VA MAX
INPUT
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
CAT I
TRIGGER LINK
RS-232
MADE IN
U.S.A.
ANALOG OUT
DIGITAL I/O
505V
MAX
V-SOURCE OUTPUT
505V PK TO CHASSIS
LO
HI INTERLOCK
!
!
FUSE
LINE
630mAT
(SB)
100 VAC
120 VAC
315mAT
(SB)
220 VAC
240 VAC
505V PK
505V PK
505V PK
IEEE-488
RS-232
Summary of Contents for 6487
Page 13: ......
Page 15: ......
Page 35: ...1 20 Getting Started Model 6487 User s Manual...
Page 47: ...2 12 Connections Model 6487 User s Manual...
Page 61: ...3 14 Measurements and Sourcing Voltage Model 6487 User s Manual...
Page 85: ...7 8 Remote Operation and Commands Model 6487 User s Manual...
Page 86: ...A Specifications...
Page 88: ...B GeneralMeasurement Considerations...
Page 94: ...C ExamplePrograms...
Page 101: ......
Page 103: ......
Page 104: ......