Series 3700 System Switch/Multimeter User's Manual
Section 6: Model 3750 Additional Information
3700S-900-01 Rev. C / July 2008
6-19
-- Write the reading/value into the appropriate step number
channel.calibration.step("" .. channel_num,1)
delay(6)
rdg=dmm.measure()
print(rdg)
channel.calibration.step("" .. channel_num,2,rdg)
dmm.range=1
channel.calibration.step("" .. channel_num,3)
delay(6)
rdg=dmm.measure()
print(rdg)
channel.calibration.step("" .. channel_num,4,rdg)
dmm.range=10
channel.calibration.step("" .. channel_num,5)
delay(6)
rdg=dmm.measure()
print(rdg)
channel.calibration.step("" .. channel_num,6,rdg)
dmm.range=1
channel.calibration.step("" .. channel_num,7)
delay(6)
rdg=dmm.measure()
print(rdg)
channel.calibration.step("" .. channel_num,8,rdg)
-- The current mode calibration follows
-- in a similar fashion
dmm.func="dccurrent"
dmm.range=.001
dmm.nplc=1
dmm.filter.count=100
dmm.filter.enable=1
channel.calibration.step("" .. channel_num,9)
delay(6)
rdg=dmm.measure()
print(rdg)
channel.calibration.step("" .. channel_num,10,rdg)
dmm.range=.1
channel.calibration.step("" .. channel_num,11)
delay(6)
rdg=dmm.measure()
print(rdg)
channel.calibration.step("" .. channel_num,12,rdg)
dmm.range=.01
channel.calibration.step("" .. channel_num,13)
delay(6)
rdg=dmm.measure()
print(rdg)
channel.calibration.step("" .. channel_num,14,rdg)
dmm.range=.1
channel.calibration.step("" .. channel_num,15)
delay(6)
rdg=dmm.measure()
print(rdg)
channel.calibration.step("" .. channel_num,16,rdg)
-- Final steps are to save & lock the calibration.
-- New calibration data is not used