Series 3700 System Switch/Multimeter User's Manual
Section 5: Basic Digital Multimeter (DMM) Operation
3700S-900-01 Rev. C / July 2008
5-79
Optimizing measurement speed
The configurations listed below assume that the multimeter is configured after
the ICL command reset().
DC voltage, DC current, and resistance
Select:
dmm.autozero=dmm.OFF
dmm.autodelay=dmm.OFF
dmm.nplc=0.0005
dmm.filter=dmm.OFF
dmm.autorange=dmm.OFF
dmm.measurecount>=1000
For resistance, assumed 2-wire ohms.
AC voltage and AC current
Select:
dmm.detectorbandwidth=300
dmm.autodelays=dmm.OFF
dmm.autozero=dmm.OFF
dmm.autorange=dmm.OFF
dmm.filter=dmm.OFF
dmm.nplc=0.0005
Temperature
Select:
dmm.transducer=dmm.TEMP_THERMOCOUPLE
dmm.opendetector=dmm.OFF
dmm.nplc=0.0005
dmm.autozero=dmm.OFF
dmm.filter=dmm.OFF
dmm.autodelay=dmm.OFF