Range
The range setting is ―remembered‖ by each measurement function. Selecting a function returns
the instrument to the last range setting for that function.
NOTE
A power cycle or an instrument reset will reset "remembered" ranges.
Measurement ranges and maximum readings
The range that is selected affects both measurement accuracy as well as the maximum
measurable level. Input values that exceed the maximum readings cause ―Overflow‖ to be
displayed.
Function
Ranges
Maximum reading
DCV (DC voltage)
100mV, 1V, 10V, 100V, 300V
± 303V
ACV (AC voltage)
100mV, 1V, 10V, 100V, 300V
± 303V
DCI (DC current)
10µA, 100µA, 1mA, 10mA,
100mA, 1A
± 3.1A
ACI (AC current)
1mA, 10mA, 100mA, 1A
± 3.1A
2 (two-wire ohm)
10
, 100
, 1k
, 10k
, 100k
,
1M
, 10M
, 100M
120M
4 (four-wire ohm)
1
, 10
, 100
, 1k
, 10k
,
100k
, 1M
, 10M
, 100M
120M
4 OC (four-wire ohm offset
compensated)
1
, 10
, 100
, 1k
, 10k
12k
4 DRY+ (four-wire ohm
dry circuit)
1
, 10
, 100
, 1k
, 10k
2.4k
TMP (temperature)
-200°C to 1820°C
Sensor dependent
FREQ (frequency)
3Hz to 500kHz
PER (period)
2µs to 333ms
Section 4
Range, Digits, Rate, Bandwidth, and Filter