Series 3700 System Switch/Multimeter Reference Manual
Section 9
:
Instrument Control Library (ICL)
Document Number: 3700S-901-01 Rev. A / August 2007
9-87
dmm.filter.count
Also see
dmm.filter.enable
(on page 9-87)
dmm.filter.type
(on page 9-87)
Example
To set the filter count for two-wire ohms to 5:
dmm.func = "twowireohms"
dmm.filter.count = 5
dmm.filter.enable
Attribute
Enables or disables filtered measurements for the selected DMM function.
Usage
To read the filter enable:
value = dmm.filter.enable
value
: Represents the present filter enable setting
To write the filter enable:
dmm.filter.enable = value
value
: Represents the desired filter enable setting. Use one of the following:
dmm.ON or 1 to enable filter measurements
dmm.OFF or 0 to disable filter measurements
Remarks
This is the filter enable setting for the DMM and it applies to the selected function as
indicated by dmm.func. Querying the setting when the selected function does not support
it will cause nil to be returned.
This attribute indicates whether filtered measurements are enabled (or not).
Changing functions with dmm.func has no effect on this setting.
The
dmm.reset
(on page 9-112) function disables the filter.
Also see
dmm.filter.count
(on page 9-86),
dmm.filter.type
(on page 9-87),
dmm.filter.window
(on
page 9-88)
Example
To enable the filter for two-wire ohms:
dmm.func = "twowireohms"
dmm.filter.enable = dmm.ON