Section 10
:
Verification
Series 3700 System Switch/Multimeter Reference Manual
10-10
Document Number: 3700S-901-01 Rev. A / August 2007
Figure 10-3: DC current verification 10uA-100uA ranges
DC current verification data 10uA-100uA ranges
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications. See
Example reading limit calculation
(on page 10-3).
Applied current
Range
Current
Lower limit
Upper limit
Verify DC curr 10
A
1.00E-05
0.00E+00 -3.000000E-10
3.000000E-10
Verify DC curr 10
A
1.00E-05
1.00E-05 9.996700E-06
1.000330E-05
Verify DC curr 10
A
1.00E-05
-1.00E-05 -1.000330E-05
-9.996700E-06
Zero 100
A
Verify DC curr 100
A
1.00E-04
1.00E-04 9.996910E-05
1.000309E-04
Verify DC curr 100
A
1.00E-04
-1.00E-04 -1.000309E-04
-9.996910E-05