Section 10
:
Verification
Series 3700 System Switch/Multimeter Reference Manual
10-20
Document Number: 3700S-901-01 Rev. A / August 2007
3. Select the SLOW integration rate with the
RATE
key.
4. Set the Series 3700 for the 1
range, and make sure the FILTER is on. Enable OC+ (offset-
compensated ohms). (Use OC+ for 1
and 10
range verification.)
5. Recalculate reading limits based on actual discrete resistor resistance values.
6. Repeat using the 10
discrete resistor on the 10
range.
Figure 10-10: Verifying discrete resistance
Discrete resistance verification data
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications (see
Calculating resistance reading limits
(on page 10-
3)).
Applied Resistance
Range
Value
Lower limit
Upper limit
Verify Res 1 Ohm
1.00E+00
1.00E+00
9.998600E-01
1.00
Verify Dry Circuit 1 Ohm
1.00E+00
1.00E+00
9.998500E-01
1.00
Verify Res 10 Ohm
1.00E+01
1.00E+01
9.00
1.01
Verify Dry Circuit 10 Ohm
1.00E+01
1.00E+01
9.00
1.01