2-20
Calibration
Model 2750 Multimeter/Switch System Service Manual
Manufacturing calibration
The manufacturing calibration procedure is normally performed only at the factory, but
the necessary steps are included here in case the unit is repaired, and the unit requires
these calibration procedures.
NOTE
If the unit has been repaired, the entire comprehensive calibration procedure
should also be performed in addition to the manufacturing calibration
procedure.
Recommended test equipment
summarizes the test equipment required for the manufacturing calibration
steps. In addition, you will need the calibrator (
) and signal generator to com-
plete the comprehensive calibration steps.
Extender board preparation
Before performing manufacturing calibration, short the output HI, LO, SHI, and SLO ter-
minals of the 7797 Calibration/Extender together using clean, solid copper wires. These
connections will form a low-thermal short necessary for the manufacturing calibration
procedure. The Calibration Extender Board should then be installed in scanner Slot #1.
Table 2-13
Recommended equipment for manufacturing calibration
Stanford Research Systems DS345 Function Generator:
1V RMS, 3Hz, ±5ppm
1V RMS, 1kHz, ±5ppm
Keithley Model 2001 or 2002 Digital Multimeter:
1V, 3Hz AC, ±0.13%
Keithley 7797 Calibration/Extender Board
Summary of Contents for 2750
Page 13: ...1 PerformanceVerification...
Page 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Page 49: ...2 Calibration...
Page 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Page 75: ...Routine Maintenance 3 RoutineMaintenance...
Page 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Page 87: ...Troubleshooting 4 Troubleshooting...
Page 111: ...Disassembly 5 Disassembly...
Page 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Page 119: ...Replaceable Parts 6 ReplaceableParts...
Page 135: ...Specifications A AccuracyandOptimization...
Page 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Page 141: ...Calibration Reference B CalibrationReference...
Page 167: ...Calibration Program C CalibrationProgram...
Page 178: ......
Page 179: ......