B-26
Calibration Reference
Model 2750 Multimeter/Switch System Service Manual
Using the *OPC command
The *OPC (operation complete) command can also be used to detect the completion of
each calibration step. To use *OPC to detect the end of each calibration step, you must do
the following:
1.
Enable operation complete by sending *ESE 1. This command sets the OPC (oper-
ation complete bit) in the standard event enable register, allowing operation com-
plete status from the standard event status register to set the ESB (event summary
bit) in the status byte when operation complete is detected.
2.
Send the *OPC command immediately following each calibration command. For
example:
:CAL:PROT:DC:STEP1;*OPC
3.
Note that you must include the semicolon (;) to separate the two commands, and
that the *OPC command must appear on the same line as the calibration command.
4.
After sending a calibration command, repeatedly test the ESB (Event Summary)
bit (bit 5) in the status byte until it is set. (Use either the *STB? query or serial poll-
ing to request the status byte.)
5.
Once operation complete has been detected, clear OPC status using one of two
methods: (1) Use the *ESR? query, then read the response to clear the standard
event status register, or (2) send the *CLS command to clear the status registers.
Note that sending *CLS will also clear the error queue and operation complete
status.
Generating an SRQ on calibration complete
An IEEE-488 bus SRQ (service request) can be used to detect operation complete instead
of repeatedly polling the Model 2750. To use this method, send both *ESE 1 and *SRE 32
to the instrument, then include the *OPC command at the end of each calibration com-
mand line, as covered above. Refer to your controller’s documentation for information on
detecting and servicing SRQs.
Summary of Contents for 2750
Page 13: ...1 PerformanceVerification...
Page 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Page 49: ...2 Calibration...
Page 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Page 75: ...Routine Maintenance 3 RoutineMaintenance...
Page 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Page 87: ...Troubleshooting 4 Troubleshooting...
Page 111: ...Disassembly 5 Disassembly...
Page 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Page 119: ...Replaceable Parts 6 ReplaceableParts...
Page 135: ...Specifications A AccuracyandOptimization...
Page 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Page 141: ...Calibration Reference B CalibrationReference...
Page 167: ...Calibration Program C CalibrationProgram...
Page 178: ......
Page 179: ......